MC-0306A Multimeter Clamp Meter Calibration Device

Model: MC-0306A Quantity: 1 set Manufacturer: Hubei MECI Electric Power Equipment Co., Ltd

Category: Metrological verification device

Keywords: MC-0306A Multimeter Clamp Meter Calibration Device

Product Overview

This instrument is a Class 0.05 (0.1) single-phase AC and DC standard device designed in accordance with the requirements of Verification Procedure JJG 124-2005, “Verification Procedure for Ammeters, Voltmeters, Wattmeters, and Ohmmeters.” The signal generation section of the source employs a DSP and a 16-bit high-speed digital-to-analog converter to produce controllable sine-wave and distorted-wave signals. The display features a 5.7-inch TFT color LCD screen, delivering crisp and vibrant visuals. The circuit design is sophisticated, ensuring reliable operation and user-friendly control.

Product Features

1. The device consists of a single-phase AC voltage standard source, a current standard source, and DC voltage and current standard sources.

2. It features a calibration function, allowing the output value of the standard source to be corrected based on the readings of an external reference standard.

3. By using a dedicated terminal block in conjunction with the specially designed “clamp meter calibration coil,” clamp meters can be tested, with a maximum test current of 1000 A.

4. Manual and semi-automatic verification of digital multimeters, analog multimeters, clamp meters, and panel-mounted digital instruments with accuracy classes of 0.2 or lower is permitted.

Including digital instruments with DC analog output.

5. It can generate distorted waveforms containing 2 to 31 harmonics, with the number of harmonics, their order, amplitude, and phase relative to the fundamental all being programmably controlled.

6. Equipped with an RS-232 interface or a USB interface and a large-capacity non-volatile memory, the device can store test data for up to 500 meters under inspection. Using host-computer software, the device can be controlled by a PC to perform both automatic and manual tests, as well as to process and manage the test results.

Technical Specifications

AC output

  • Voltage ranges: 200 mV, 400 mV, 2000 mV, 4000 mV, 20 V, 100 V, 200 V, 500 V, 1000 V;
  • Current ranges: 2 mA, 4 mA, 20 mA, 40 mA, 0.2 A, 0.5 A, 2 A, 5 A, 10 A, 20 A;
  • Adjustment range: 0–120% (except for 1000 V);
  • Fineness adjustment: 5×10⁻⁵;
  • Voltage and current output capacity: maximum 20 VA;
  • Voltage and current setting accuracy

A: Voltage range: 20–1000 V; current range: 0.2–20 A; accuracy: ±(0.03% of reading + 0.02% of full scale).

B: Voltage range: 200–4000 mV; current range: 2–40 mA; accuracy: ±(0.3% of reading + 0.2% of full scale).

  • Voltage and current output stability: A: ≤0.01%/60 s; B: ≤0.05%/60 s;
  • Voltage and current waveform distortion ≤0.3%;
  • Sine wave output frequency adjustment range: (45–65)·N Hz (N = 1, 2, 3, …, 31);
  • Sine-wave output frequency resolution: 0.001·N Hz (N = 1, 2, 3, …, 31);
  • Sine wave output frequency adjustment accuracy: 0.01 Hz;
  • Harmonic output from the 2nd to the 31st harmonic, with amplitude ranging from 0% to 20% of the fundamental rated value; phase resolution of 0.01°·N (where N = 2, 3, 4, …, 31); the output can simultaneously contain multiple harmonics, or the fundamental amplitude can be set to zero, resulting in an output that is a sinusoidal waveform at a frequency higher than the power frequency.

 

DC output

  • Voltage ranges: 200 mV, 400 mV, 2000 mV, 4000 mV, 20 V, 40 V, 100 V, 200 V, 400 V, 1000 V;
  • Current ranges: 20 μA, 40 μA, 200 μA, 400 μA, 2 mA, 4 mA, 20 mA, 40 mA, 0.2 A, 1 A, 2.5 A, 5 A, 10 A, 20 A;
  • Voltage and current adjustment range: 0–110% (except for 1000 V).
  • Voltage and current adjustment resolution: 5×10⁻⁵;
  • Voltage and current output capacity: maximum 20 W;
  • Voltage setting accuracy: ±(0.03%RD + 0.02%RG);
  • Current setpoint accuracy: ±(0.06%RD + 0.04%RG);
  • Voltage and current output stability ≤0.01%/60s;
  • Voltage and current output ripple ≤1%.

 

Frequency Output

  • Sine wave output range (AC voltage/current supply): (45–65)·N Hz;
  • Resolution: 0.001·N Hz (N = 1, 2, 3, …, 31);
  • Adjustment accuracy: 5×10⁻⁵.

 

Single-frequency square wave

  • Single-frequency square-wave outputs at 1 kHz, 5 kHz, 10 kHz, 50 kHz, 100 kHz, 200 kHz, and 1000 kHz;
  • Square wave amplitude: 4.5 V;
  • Adjustment accuracy: 5×10⁻⁵.

 

Electronic resistance box output

  • Measurement ranges: 0–200 Ω, 0–2 kΩ, 0–20 kΩ, 0–200 kΩ, 0–2 MΩ;
  • Fineness adjustment: 0.025%;
  • Accuracy for the 0–200 Ω and 0–2 MΩ ranges: ±(0.3% RD + 0.2% RG); for all other ranges: ±(0.12% RD + 0.08% RG).

 

Power supply: single-phase, 220 V ±10%, 50 Hz ±5%.

Operating environment: temperature 20°C ± 10°C, relative humidity ≤85% RH.

Dimensions: 440 × 460 × 165 mm; weight: approximately 12 kg.